Abstract:
Transcutaneous oximetry measurement (TCOM) is a non-invasive method of determining oxygen tension at the skin level using heated electrodes. To compare TCOM values generated by different machines and to establish lower limb TCOM values in a cohort of healthy individuals younger than 40 years of age. Sixteen healthy, non-smoking volunteers aged 18 to 39 years were recruited. TCOM was obtained at six locations on the lower leg and foot using three different Radiometer machines. Measurements were taken with subjects lying supine, breathing air. Except for one sensor site, there were no statistical differences in measurements obtained by the different TCOM machines. There was no statistical difference in measurements comparing left and right legs. Room air TCOM values for the different lower leg sites were (mean (SD) in mmHg): lateral leg 61.5 (9.2); lateral ankle 61.1 (9.7); medial ankle 59.1 (10.8); foot, first and second toe 63.4 (10.6); foot, fifth toe 59.9 (13.2) and plantar foot 74.1 (8.8). The overall mean TCOM value for the lower limb was 61 (10.8; 95% confidence intervals 60.05-62.0) mmHg.
Trinks, Blake, Young, Thistlethwaite, Vangaveti, , , , (2017). Transcutaneous oximetry measurements of the leg: comparing different measuring equipment and establishing values in healthy young adults. Diving and hyperbaric medicine, 2017 Jun;47(2):82-87. https://www.ncbi.nlm.nih.gov/pubmed/28641320